Notes
Slide Show
Outline
1
CT Scanner Workshop
2
CT Scanner Workshop
  • Lastditch General Hospital has at last been given funding to replace its outdated single slice CT scanner, a Reliatom Basic.
  • As they wish to undertake cardiac scanning, their business case for a  64 slice model has been approved.
  • Of the four 64-slice models offered, they have shortlisted the following two:
    • The Superwhiz TopSpin 64
    • The Ultra Volume Excel 64
  • Specifications of the two new scanners are given


3
CT Scanner Workshop
  • Leaving aside financial issues, consider the pros and cons of the two scanners in terms of
    • scanning capabilities (eg coverage  - speed and length)
    • image quality (resolution)
    • patient dose.
  • There is not necessarily any right answer, and you may think some information is irrelevant
  • The aim is to stimulate discussion between delegates with different professional background and experience
  • Same groups as for quiz
4
Disclaimer
  • Whilst this is based on true life, any resemblances to current or redundant scanner models is purely co-incidental!
5
Academic Exercise
  • In reality would not look at some of these specifications in so much detail
    • Use it to back up performance data
  • A suggestion to go through this is to
    • 1. first find those aspects that you think are irrelevant
    • 2. Since a number of the specifications are inter-related, identify those which relate to each of the topics (coverage, spatial resolution (scan plane and z), and dose). Then discuss.
  • An alternative is to go through each specification one at a time.
6
1. Gantry Aperture
  • UVE64 larger.
    • Good for patient access for interventional work.
    • Potential for RT treatment planning and bariatric patients.
  • Would like also to know what largest field of view is, and image quality for largest field of view
7
2. Maximum Field of View
  • UVE64 is larger.
    • Good for RT planning and bariatric patients
    • Might want to find out image quality of largest field of view.
    • Can be reduced outside of main fov (Still suitable for identifying patient contour)
8
3. Rotation Times
  • ST64 Shortest minimum scan time
    • Less movement artefact, more contrast phases….
    • Short rotation  time important in cardiac as determines temporal resolution
    • A variety of times can be helpful for optimum cardiac scanning at all heart rates
    • Make sure that the minimum scan time is for 360o
    • Don’t forget when operating at a fast scan time, tube will be operating at higher mA values (good heat capacity)
9
4. Minimum Slice Width
  • ST 64, thinner slice ie better z-axis resolution
    •  Don’t just believe on face value if this is important to you. Look at measured data, and or rounding.
    • Eg if image width is quoted to nearest 0.5 mm then both could be the same…..
10
5. Maximum number of simultaneous slices
  • ‘image slices’ or ‘data slices’ ?
    • ST may acquire 64 channels of data using the z-flying focal spot, earning it the right to be called a 64 (data) slice scanner
    • Z-resolution potentially better – but not necessarily as interleaved sampling from complementary projections and low pitches can give a similar results (see item 13.)
11
6. Total length of detector array
12
7. Number of detectors per row
  • Affects spatial resolution in scan plane
13
7. Number of detectors per row
14
Scan plane (x-y) Resolution
  • Determined by
    • Effective detector size at iso-centre, and this is dependent on
      • No detector elements per row, geometry (FID, FDD)
      • Focal spot size
    • Number of samples or projections per rotation
    • See also items 9 and 11
  • If assume same maximum field of view for full image quality (500). Then effective detector width for ST = 500/600, and for UVE =  = 500/800. Therefore UVE potentially better spatial resolution.


15
Effective Detector Width
  • Fundamental limitation of spatial resolution
  • Depends on
    • Focal spot size
    • Scanner geometry (FID, FDD)
    • Detector element size
16
7. Number of detectors per row
17
8. Detector Type
  • Both solid state ceramic but nothing to say which is best
    •  afterglow
    •  ring artefacts
    •  detector efficiency
18
9. Focal spot size
  • Affects resolution (x-y, and z)
  • UVE64 smaller fs size. All things being equal
    • Improved scan plane resolution
    • Smaller penumbra - lower dose
19
10. Focus to Detector Distance
  • UVE64 (110) vs ST64 (90)
    • UVE64 Requires higher mAs for same noise (and dose)
      • tube heat capacity and dissipation requirements important
20
10. Focus to Detector Distance
  • UVE64 (110) vs ST64 (90)
    • More room within gantry so collimators can be further away from focus-smaller penumbra
    • ?Less cone bean artefact as beam in z-direction more parallel
    • Potentially less scatter.
21
11. Sampling Rate
  • Scan plane resolution depends on sampling interval
  • Note ST64 is 2500 samples/sec
    • At 0.3 sec, this is 750 samples/rotation
    • At 0.4s, 1000samples/rotation. Neither as good as UVE
22
Sampling Rate
  • Sampling rate and scan time determine number of projections per image
  • Potential limit on spatial resolution
23
Scan plane (x-y) Resolution
  •    Spatial Resolution
24
Scan plane (x-y) Resolution
25
Scan plane (x-y) Resolution
26
12. Flying (Dynamic) Focal Spot (x-y)
  • UVE 64 - samples at half detector width (in x-y). Potential for improved scan plane resolution,
27
13. Flying (Dynamic) Focal Spot (z)
  • ST64 – samples at half detector width in z-axis. Potential for improved z-axis resolution
28
13. Flying (Dynamic) Focal Spot (z)
  • ‘image slices’ or ‘data slices’ ?
    • ST may acquire 64 channels of data using the z-flying focal spot, earning it the right to be called a 64 (data) slice scanner
    • Z-resolution potentially better – but not necessarily as interleaved sampling from complementary projections and low pitch can give a similar effect.
29
 Z-Axis flying focal spot vs interleaved samples
  • Z-Axis Flying Focal Spot
    • To improve spatial resolution in helical scanning
30
 Z-Axis flying focal spot vs interleaved samples
  • Z-axis interpolation
    • To improve z-axis resolution in helical scanning
31
14. Helical Interpolation
  • ST64. Potentially more exact, potentially takes longer. But don’t know which is best
32
15, 16, 17 X-ray tube
  • 15. Anode Cooling
    • UVE64 potentially much greater cooling efficiency as shown in next point
  • 16. Anode heat capacity
    • UVE64
    • Longer scan lengths possible
    •  Higher mAs possible
    • Could lead to high doses if used carelessly
  • 17. Anode cooling rate


33
18. Filtration
  • UVE64 Harder beam
    •  more uniform dose distribution
    •  better dose:noise ratio
  • ST Softer beam
    •  ? better low contrast differentiation
    •  ? higher surface dose
34
19. Maximum mA
  • Good for bariatrics
  • Good for short scan times
35
20. Maximum Continuous scan time
  • ST better, but seems inconsistent with heat capacity. Question whether  data is correct?
  • Maybe limited because longer scan time would not be needed in practice?
36
21. CTDIvol
  • Note mGy/100mAs
  • ST64 has shorter fid, and lower filtration so is likely to operate at lower mA for a clinical scan (although it has a much higher max mA available)
37
21. CTDIvol
38
22. Dose modulation
  • ST does not allow rotational modulation.
    • Does not have benefit of streaking artefact reduction
    • Does not make full use of possibility for dose reduction
39
CT Scanner Workshop