|
1
|
|
|
2
|
- Lastditch General Hospital has at last been given funding to replace its
outdated single slice CT scanner, a Reliatom Basic.
- As they wish to undertake cardiac scanning, their business case for
a 64 slice model has been
approved.
- Of the four 64-slice models offered, they have shortlisted the following
two:
- The Superwhiz TopSpin 64
- The Ultra Volume Excel 64
- Specifications of the two new scanners are given
|
|
3
|
- Leaving aside financial issues, consider the pros and cons of the two
scanners in terms of
- scanning capabilities (eg coverage
- speed and length)
- image quality (resolution)
- patient dose.
- There is not necessarily any right answer, and you may think some
information is irrelevant
- The aim is to stimulate discussion between delegates with different
professional background and experience
- Same groups as for quiz
|
|
4
|
- Whilst this is based on true life, any resemblances to current or
redundant scanner models is purely co-incidental!
|
|
5
|
- In reality would not look at some of these specifications in so much
detail
- Use it to back up performance data
- A suggestion to go through this is to
- 1. first find those aspects that you think are irrelevant
- 2. Since a number of the specifications are inter-related, identify
those which relate to each of the topics (coverage, spatial resolution
(scan plane and z), and dose). Then discuss.
- An alternative is to go through each specification one at a time.
|
|
6
|
- UVE64 larger.
- Good for patient access for interventional work.
- Potential for RT treatment planning and bariatric patients.
- Would like also to know what largest field of view is, and image quality
for largest field of view
|
|
7
|
- UVE64 is larger.
- Good for RT planning and bariatric patients
- Might want to find out image quality of largest field of view.
- Can be reduced outside of main fov (Still suitable for identifying
patient contour)
|
|
8
|
- ST64 Shortest minimum scan time
- Less movement artefact, more contrast phases….
- Short rotation time important in
cardiac as determines temporal resolution
- A variety of times can be helpful for optimum cardiac scanning at all
heart rates
- Make sure that the minimum scan time is for 360o
- Don’t forget when operating at a fast scan time, tube will be operating
at higher mA values (good heat capacity)
|
|
9
|
- ST 64, thinner slice ie better z-axis resolution
- Don’t just believe on face value
if this is important to you. Look at measured data, and or rounding.
- Eg if image width is quoted to nearest 0.5 mm then both could be the
same…..
|
|
10
|
- ‘image slices’ or ‘data slices’ ?
- ST may acquire 64 channels of data using the z-flying focal spot,
earning it the right to be called a 64 (data) slice scanner
- Z-resolution potentially better – but not necessarily as interleaved
sampling from complementary projections and low pitches can give a
similar results (see item 13.)
|
|
11
|
|
|
12
|
- Affects spatial resolution in scan plane
|
|
13
|
|
|
14
|
- Determined by
- Effective detector size at iso-centre, and this is dependent on
- No detector elements per row, geometry (FID, FDD)
- Focal spot size
- Number of samples or projections per rotation
- See also items 9 and 11
- If assume same maximum field of view for full image quality (500). Then
effective detector width for ST = 500/600, and for UVE = = 500/800. Therefore UVE potentially
better spatial resolution.
|
|
15
|
- Fundamental limitation of spatial resolution
- Depends on
- Focal spot size
- Scanner geometry (FID, FDD)
- Detector element size
|
|
16
|
|
|
17
|
- Both solid state ceramic but nothing to say which is best
- afterglow
- ring artefacts
- detector efficiency
|
|
18
|
- Affects resolution (x-y, and z)
- UVE64 smaller fs size. All things being equal
- Improved scan plane resolution
- Smaller penumbra - lower dose
|
|
19
|
- UVE64 (110) vs ST64 (90)
- UVE64 Requires higher mAs for same noise (and dose)
- tube heat capacity and dissipation requirements important
|
|
20
|
- UVE64 (110) vs ST64 (90)
- More room within gantry so collimators can be further away from
focus-smaller penumbra
- ?Less cone bean artefact as beam in z-direction more parallel
- Potentially less scatter.
|
|
21
|
- Scan plane resolution depends on sampling interval
- Note ST64 is 2500 samples/sec
- At 0.3 sec, this is 750 samples/rotation
- At 0.4s, 1000samples/rotation. Neither as good as UVE
|
|
22
|
- Sampling rate and scan time determine number of projections per image
- Potential limit on spatial resolution
|
|
23
|
|
|
24
|
|
|
25
|
|
|
26
|
- UVE 64 - samples at half detector width (in x-y). Potential for improved
scan plane resolution,
|
|
27
|
- ST64 – samples at half detector width in z-axis. Potential for improved
z-axis resolution
|
|
28
|
- ‘image slices’ or ‘data slices’ ?
- ST may acquire 64 channels of data using the z-flying focal spot,
earning it the right to be called a 64 (data) slice scanner
- Z-resolution potentially better – but not necessarily as interleaved
sampling from complementary projections and low pitch can give a
similar effect.
|
|
29
|
- Z-Axis Flying Focal Spot
- To improve spatial resolution in helical scanning
|
|
30
|
- Z-axis interpolation
- To improve z-axis resolution in helical scanning
|
|
31
|
- ST64. Potentially more exact, potentially takes longer. But don’t know
which is best
|
|
32
|
- 15. Anode Cooling
- UVE64 potentially much greater cooling efficiency as shown in next
point
- 16. Anode heat capacity
- UVE64
- Longer scan lengths possible
- Higher mAs possible
- Could lead to high doses if used carelessly
- 17. Anode cooling rate
|
|
33
|
- UVE64 Harder beam
- more uniform dose distribution
- better dose:noise ratio
- ST Softer beam
- ? better low contrast
differentiation
- ? higher surface dose
|
|
34
|
- Good for bariatrics
- Good for short scan times
|
|
35
|
- ST better, but seems inconsistent with heat capacity. Question
whether data is correct?
- Maybe limited because longer scan time would not be needed in practice?
|
|
36
|
- Note mGy/100mAs
- ST64 has shorter fid, and lower filtration so is likely to operate at
lower mA for a clinical scan (although it has a much higher max mA
available)
|
|
37
|
|
|
38
|
- ST does not allow rotational modulation.
- Does not have benefit of streaking artefact reduction
- Does not make full use of possibility for dose reduction
|
|
39
|
|